|
information
on:
|
Water analysis
Guidelines for the determination of total organic carbon (TOC) and dissolved organic
carbon (DOC)
ISO-CEN EN 1484
This European standard was approved
by CEN on 1997-04-06. CEN members are bound to comply with the CEN/CENELEC Internal
Regulations which stipulate the conditions for giving this European Standard the status of
a national standard without any alteration.
Up-to-date lists and bibliographical references
concerning such national standards may be obtained on application to the Central
Secretariat or to any CEN member.
The European standards exist in three official versions (English, French, German). A
version in any other language made by translation under the responsibility of a CEN member
into its own language and notified to the Central Secretariat has the same status as the
official version.
CEN members are the national standards bodies of Austria, Belgium,, Czech Republic,
Denmark, Finland, Germany, Greece, Iceland, Ireland, Italy, Luxembourg, Netherlands,
Norway, Portugal, Spain, Sweden, Switzerland and United Kingdom.
Foreword
This European Standard has been prepared by the Technical Committee CEN/TC 230 "Water
Analysis", the secretariat of which is held by DIN.
This European Standard contains three informative annexes.
This European Standard shall be given he status of a national standard, either by
publication of an identical test or by endorsement at the latest by November 1997, and
conflicting national standards shall be withdrawn at the latest by November 1997.
According to the CEN/CENELEC Internal Regulations, the national standards organizations of
the following countries are bound to implement this European Standard: Austria, Belgium,
Czech Republic, Denmark, Finland, France, Germany, Greece, Iceland, Ireland, Italy,
Luxembourg, Netherlands, Norway, Portugal, Spain, Sweden, Switzerland and the United
Kingdom.
Introduction
It is absolutely essential that tests conducted according to
this standard are carried out by suitably qualified staff.
Total organic carbon (TOC) is a measure of the carbon
content of dissolved and undissolved organic matter present in the water. It does not give
information on the nature of the organic substance.
1. Scope
This European Standard gives guidance for the determination of organic carbon in drinking
water, ground water, surface water, sea water and waste water. It deals with definitions,
interferences, reagents, and the same pretreatment of water samples with a content of
organic carbon ranging from 0,3 mg/l to 1000 mg/l while the lower value is only applicable
in special cases, for example drinking water, measured with instruments capable of
measuring these low levels. Higher concentrations may be determined after appropriate
dilution. This European Standard does not deal with the instrument-dependent aspects.
In addition to organic carbon the water sample may contain carbon dioxide or ions of
carbonic acid. Prior to the TOC determination, it is essential that this inorganic carbon
is removed by purging the acidified sample with a gas which is free from CO2 and organic
compounds. Alternatively, both total carbon (TC) and total inorganic carbon (TIC) may be
determined and the organic carbon content (TOC) may be calculated by subtracting the total
inorganic carbon from the TC. This method is particularly suitable for samples in which
the total inorganic carbon is less that the TOC.
Purgeable organic substances, such as benzene, toluene, cyclohexane, and chloroform may
partly escape upon stripping. In presence of these substances the TOC concentration is
determined separately, or the differential method (TC - TIC = TOC) may be applied. By
using the differential method, the value of the TOC should be higher than the TIC, or at
least of similar size.
Cyanide, cyanate, and particles of elemental carbon (soot) when present, will be
determined together with the organic carbon.
NOTE: In presence of humic material low values may occur when UV- radiation is used.
2. Normative References
This European Standard incorporates by dated or
undated reference provisions from other publications. These normative references are cited
at the appropriate places in the text and the publications are listed hereafter. For dated
reference, subsequent amendment to or revisions of any of these publications apply to this
European standard only when incorporated in it by amendment or revision. For updated
reference the latest edition of the publication referred to applies-
EN ISO 5667-3 : 1995
Water Quality - Sampling - Part 3: Guidance on the preservation and handling of samples
(ISO 5667-3: 1994)
3. Definitions
For the purposes of this European Standard the
definitions apply:
3.1 Total Carbon (TC)
The sum of organically bound and inorganically
bound carbon present in water, including elemental carbon.
3.2 Total inorganic carbon (TIC)
The sum of carbon present in water, consisting of elemental carbon, total carbon dioxide,
carbon monoxide, cyanate, and thiocyanate. TOC instruments mostly register as TIC only the
CO2 originating from hydrogen carbonates and carbonates.
3.3. Total organic carbon (TOC)
The sum of organically bound carbon present in water, bonded to dissolved or suspended
matter. Cyanate, elemental carbon and thiocyanate will also be measured.
3.4 Dissolved organic carbon (DOC)
The sum of organically bound carbon present in water originating from compounds which will
pass a membrane filter of pore size of 0,45 m m. Cyanate and thiocyanate are also
measured.
3.5 Volatile organic carbon (VOC, POC)
Under the conditions of this method purgeable organic matter (POC).
3.6 Non volatile organic carbon (NVOC,
NPOC)
Under the conditions of this method non-purgeable
organic carbon (NPOC).
4. Principle
Oxidation of organic carbon (org. C) in water to
carbon dioxide by combustion, by the addition of an appropriate oxidant, by UV radiation
or any other high-energy radiation.
The application of the ultraviolet method with only oxygen gas as an oxidant is restricted
to low polluted waters containing low concentrations of TOC.
Inorganic carbon is removed by acidification and purging or is determined separately.
The carbon dioxide formed by oxidation is determined either directly or after reduction,
for example, to methane (CH4).
The final determination of CO2 is carried out by a number of different procedures, for
example: Infrared spectrometry, titration (preferably in non-aqueous solution), thermal
conductivity, conductometry, coulometry, CO2-sensitive sensors and flame ionization
detection - used after reduction of the CO2, among others, to methane.
5. Reagents
5.1 General remarks
Use only reagents of recognized analytical grade.
In this European Standard only those chemicals and gasses are listed which are used with
the majority of TOC methods. Reagents should be used according to the manufacturers
instructions, and should be pretreated, if necessary.
The TOC of the water being used for dilution and for preparation of the calibration
standards should be negligibly low, compared with the lowest TOC concentration to be
determined.
The method for water treatment depends on the concentration range under investigation as
shown in table 1.
NOTE: For measurements of a TOC concentration< 0,5 mg/l, it is preferable to prepare
water for blanks and the calibration solutions immediately prior to analysis (see table
1).
Table 1: Dilution Water Specifications
TOC
of the Sample (mg/l C) |
Maximum
acceptable TOC of the dilution water (mg/l C) |
Example
of water treatment methods |
< 10 |
0,1*
0,3 |
UV treatment
condensation |
10 to 100 |
0,5 |
double distillation
with KMnO4/K2Cr2O7 |
> 100 |
1 |
Distillation |
*only for ultrapure
water |
|
|
5.2 Potassium hydrogen phthalate, stock solution, p
(org. C) = 1 000 mg/l
Dissolve in a 1 000 ml one-mark volumetric flask 2,125 g of potassium hydrogen phthalate
(C8H5KO4) (dried for 1 h at a temperature between 105°C and 120°C) in 700 ml of water,
and make up to volume with water.
The solution is stable for about 2 months if stored in a tightly stoppered bottle in a
refrigerator.
5.3 Potassium hydrogen phthalate, standard
solution, p (org.C) = 100 mg/l
Pipette 100 ml of the potassium hydrogen phthalate stock solution (5.2) into a 1 000 ml
one-mark volumetric flask, and make up to volume with water.
The solution is stable for about 1 week if stored in tightly stoppered bottle in a
refrigerator.
5.4 Standard Solution for the determination of
inorganic carbon, p(inorg.C) = 1 000 mg/l
Dissolve in a 1 000 ml one-mark volumetric flask 4,415 g of sodium carbonate (Na2CO3),
dried for 1 h at (285± 5)°C in approximately 500 ml of water.
Add 3,500 g of sodium hydrogen carbonate (NaHCO3) (dried for 2 h over silica gel), and
make up to volume with water.
This solution is stable at room temperature for about 2 weeks.
5.5 Substances , not readily oxidizable
A standard solution shall be used to check the operational performance of the system-
NOTE: In the interlaboratory trial, copper phthalocyanine has been used for this purpose.
A suitable test solution of copper phthalocyanine, p(org.C) = 100 mg/l, may be
prepared as follows: In a 1 000 ml volumetric flask 0,256 g of copper
phthalocyanine-tetrasulfonic acid (tetrasodium salt) (C32H12CuN8O12S4Na4) are dissolved in
700 ml of water, and made up to volume with water. The solution is stable for about 2
weeks. WARNING: This reagent is toxic.
5.6 Further reagents
5.6.1 The reagents 5.2, 5.4 and 5.5 may be replaced by other reagents provided these are
stable titrimetric substances.
5.6.2 Non volatile acids for expelling the carbon dioxide, such as phosphoric acid
c(H3PO4) = 0,5 mol/l, if necessary, more concentrated.
5.7 Gases
Air, nitrogen, oxygen, free from carbon dioxide and organic impurities. Use other gases in
accordance with the instrument manufacturers specifications.
6. Apparatus
Usual laboratory apparatus and
6.1 Apparatus for the determination of organic carbon
6.2 Homogenization device, for example, a magnetic stirrer with adequate performance for
the homogenization of dispersed matter, a suitable ultrasonic apparatus or a high speed
stirrer.
7. Sampling and samples
7.1 Sampling
See also EN ISO 5667-3.
When sampling ensure that the samples being collected are representative (particularly in
the presence of undissolved substances), and take care not to contaminate the samples with
organic substances.
Collect water samples in glass or polyethylene bottles, completely filled with the sample,
and, if biological activity is suspected, acidify to pH 2 (for example with phosphoric
acid (5.6.2)). In some cases the result may be reduced by the loss of volatile substances
during to the liberation of carbon dioxide, when the sample is acidified. If volatile
organic compounds are suspected, carry out the measurement without acidification and
within 8 h of sampling. Otherwise, store the sample in a refrigerator at a temperature in
the range of 2 °C to 5 °C and analyze within 7 days. If this is not possible, the sample
can be kept at - 15° C to - 20°C for up to several weeks.
7.2 Preparation of the water sample
If, because of the inhomogenity , the collection of a representative sample is not
possible, even after thorough shaking, homogenize the sample, using for example, the
apparatus as described in 6.2.
The homogeneity may be checked for example by analyzing a sample from the upper and the
lower layer of the bottle.
If only dissolved organic substances (DOC) are to be determined, filter the sample through
a membrane filter, pore size 0,45 m m, which has previously been washed with hot
water to completely remove adhering organic substances. Nevertheless the carbon content of
the filtrate has to be determined and taken into account.
8. Procedure
8.1 Calibration
Comparison methods (for example IR detection) require calibration; in the case of absolute
methods, for example acidimetry or coulometry, the calibration serves the purpose of
checking the analytical system.
Calibrate the instrument according to the manufacturers instructions.
Establish a calibration curve by analyzing potassium hydrogen phthalate standard solutions
of adequate concentrations. For example for mass concentrations ranging from 10 mg/l to
100 mg/l proceed as follows: Prepare a series of at least 5 calibration solutions from the
potassium hydrogen phthalate stock solution (5.2), and make up to volume with water.
Analyze each solution and the blank solution (from the flask without added potassium
hydrogen phthalate) in accordance with the manufacturers instructions.
Establish a calibration curve by plotting the mass concentrations of TOC, in milligrams
per litre of carbon, against the instrument-specific response units (l).
The reciprocal value of the slope of the resulting calibration line is the calibration
factor f, in milligrams per litre of carbon.
In the case of the determination of TIC it is necessary to establish a calibration curve
by analyzing calibration solutions made from solution (5.4).
In the case of the determination of TOC from the difference TC-TIC, it is necessary to
establish a calibration curve by analyzing calibration solutions prepared from a known
mixture of the standard solutions 5.3 and 5.4.
8.2 Control Procedures
Analyze the test solutions (either 5.2; or 5.3; or 5.5; or 5.6.1) with each series of
samples in order to verify the accuracy of the results obtained by the method.
If the deviations found are higher than intra-laboratory quality criteria allow, then the
following causes of error should be investigated.:
- Malfunction of the instrument (for example in the oxidation or detection system,
leakages, faults in the temperature or the gas control);
- Changes in concentration in the test solution;
- Contamination of the measuring assembly.
Regularly check the entire measuring system according to the instructions of the
manufacturer; also check the entire systems regularly for the absence of leaks.
These control experiments are carried out in addition to the instrumental control as
specified in the operating instructions provided by the instrument manufacturer.
8.3 Determination
Determine the TOC concentrations of the samples in accordance with the instrument
manufacturers instructions.
In the case of direct determination of the TOC, remove the total inorganic carbon (ensure
that the pH is below 2) prior to analysis. Carefully minimize the loss of volatile organic
substances.
The TOC concentration should be within the working range of the calibration. This can be
achieved by diluting the sample.
Prior to each batch of TOC determinations (for example 10 determinations) carry out
appropriate control experiments at the intervals recommended by the manufacturer or
specified by the laboratory.
After acidification, blow a stream of pure inert gas free from CO2 and organic impurities
through the system (for approximately 5 min) in order to remove CO2.
9. Evaluation of results
9.1 Calculation
Depending on the type of TOC instrument used, different kinds of readings may be obtained
from which the TOC or DOC concentration of the analyzed sample is calculated. In the case
of discontinuous measurements, these values may be in terms of, for example, peak heights,
peak areas or volume of the titrant required. Normally, peak areas are reported. Use peak
heights only if they are proportional to the concentration.
In the case of quasi-continuous TOC or DOC measurements, the CO2 concentration generated
by the combustion of the organic matter is recorded, for example as a line on a strip
chart recorder. The distance of this line from the zero line is proportional to the TOC
concentration
Calculate the mass concentration using the calibration curve (8.1).
The mass concentration of TOC or DOC, calculated in mg/l, may also be obtained from the
following equation:
I · f · v
Vp
Where:
I is the instrument specific response;
f is the calibration factor evaluated as in 8.1 in milligrams per litre of carbon;
V is the volume of the diluted water sample, in millilitres;
Vp is the sample volume being diluted to V, in millilitres.
9.2 Expression of Results
The results are expressed in milligrams per litre of carbon. The way of expressing the
results depends on the random error (precision) of the measurement. 2 or 3 significant
figures are reported.
Examples:
p(TOC )= 0,76 mg/l carbon or
p(TOC) = 530 mg/l carbon or
p(TOC) = 530 mg/l carbon or
Information on repeatability and reproducibility, as obtained from an interlaboratory
experiment, is given in the annex.
10. Test report
The test report shall contain the following
information.
a) a reference to this European Standard;
b) all data required for a complete identification of the sample;
c) details concerning the storage of the laboratory samples before analysis; including the
time between sampling and analysis;
d) sample pretreatment (time for settlement, filtration);
e) the mass concentration of TOC and DOC respectively in the sample, in milligrams per
litre;
f) details of any deviation from the procedure specified in this European Standard or any
other circumstances that may have influenced the result.
Annex A (informative)
[ 1] Dürr, W. and Merz, W., Evaluation of the TOC
- ISO interlaboratory trial and discussion of the results, Vom Wasser, 55, 287-294
(1980).
[2] Methods for the examination of waters and associated materials. The instrumental
determination of total organic carbon, total oxygen demand and related determinands.
Her Majestys Stationery Office, London, 1995.
[3] EBV (WI: 00230055) Guide on Analytical quality control (AQC) in water analysis
Annex B (informative)
Table B.1: Results of an interlaboratory
trial on the TOC determination
Sample |
Nominal value |
Total
mean |
Recovery |
Reproducibility |
Repeatability |
Number
of laboratories |
Number
of analytical results after elimination of outliers |
Outliers
of results |
|
mg/l |
mg/l |
% |
Standard deviation
mg/l |
Variation
coefficient % |
Standard deviation
mg/l |
Variation
coefficient % |
|
|
|
1 |
2,3 |
2,99 |
129,9 |
0,687 |
23 |
0,19 |
6,3 |
55 |
259 |
13 |
2 |
18,5 |
19,2 |
103,9 |
1,23 |
6,4 |
0,38 |
2 |
56 |
260 |
9 |
3 |
120 |
139 |
115,9 |
12,4 |
8,9 |
2,8 |
2 |
54 |
236 |
16 |
4 |
|
307 |
|
13,9 |
4,5 |
3,8 |
1,2 |
54 |
244 |
20 |
Sample 1= |
Copper
phthalocyaninetetrasulfonic acid, tetrasodium salt |
Sample 2= |
Mixture
of potassium hydrogen phthalate and Copper
phthalocyanine tetrasulfonic acid, tetrasodium salt |
Sample 3= |
Mixture
of potassium hydrogen carbonate Potassium hydrogen
phthalate and
Copper phthalocyanine tetrasulfonic acid, tetrasodium salt |
Sample 4= |
Industrial
waste water, real sample, filtered |
Note: The reason for the recovery rate of 130% as stated for sample 1
(sample with low TOC concentration) is possibly due to systematic errors
(non-consideration or only partial consideration of the TOC concentration of the blank
water).
The increased recovery rate in case of sample 3 is possibly
caused by the very high TIC concentration. In similar cases, the statements from the
instrument manufacturers concerning acid volume and stripping time are often non
sufficient. |
Annex C
(informative)
Determination of samples containing particles
C.1 Additional conditions
The instrumental specifications for TOC measurement should at least be suitable for
measuring particles of a size of 100 micron (convention).
NOTE 1: In the interlaboratory trial (see table
C.1), samples containing particles up to 100 micron have been measured.
NOTE 2: If the TOC determination of the sample containing particles does not lead to
reproducible results, even after intensive homogenization, the sample may be filtered and
the TOC of the filtrate and the residue may be determined separately.
Systems based on oxidation using UV are not suitable for the determination of
microcellulose as an example for samples containing suspended matter (see table C.1
results of the interlaboratory trial, Sample1 b).
C.2 Suspension to test the particle processing capability
This suspension serves to check the homogenization and the recovery of incompletely
dissolved sample components (particulate matter):
To prepare this suspension, 225 mg of cellulose, (C6H10O5)n, (particle size about 20 m m
to 100 m m, concentration of the suspension test sample: 100 mg/l C) are placed into a 1
000 ml volumetric flask, moist with water, made up to volume with water, and stirred with
a magnetic stirrer until the suspension is homogeneous. Ultrasonic treatment should not be
used because it is necessary to stir each time before use.
Suspensions of equally well suited substances may also be used as standards, for example
for the analysis of paper mill effluents.
C.3 Control of the homogenization and
recovery of incompletely dissolved sample components (particle processing capability)
It is advisable with each batch of analyses of samples containing solids to check the
homogenization and the recovery of suspended sample components (particle processing
capability of the instrument) by using a test suspension (C.2). Withdrawal of an aliquot
is to be made by stirring. If an autosampler is used, the aliquots in their vessels should
be stirred. The mean value from a triple measurement should be between 90 mg/l and 100
mg/l, the repeatability variation coefficient should be < 10 %.
NOTE 1: For this test the particle size is
important.
NOTE 2: An optimal homogenization without segregation of particles is provided, for
example, by an oscillating stirrer.
C.4 Performance Data
Table C.1: Results of an interlaboratory trial on the TOC determination
Sample |
Nominal value |
Total mean |
Recovery |
Reproducibility |
Repeatability |
Number of
laboratories |
Number
of analytical results after elimination of outliers |
Outliers
of results |
|
mg/l |
mg/l |
% |
Standard deviation
mg/l |
Variation
coefficient % |
Standard deviation
mg/l |
Variation
coefficient % |
|
|
|
1 a |
20 |
16,65 |
83,2 |
7,5 |
45,1 |
2 |
12,0 |
32 |
149 |
0 |
1 b |
20 |
0,53 |
2,7 |
0,4 |
75,0 |
0,15 |
27,3 |
15 |
56 |
10 |
sample 1a= |
Cellulose
microcrystals (combustion determination) |
sample 1b= |
Cellulose
microcrystals (UV-oxidation) |
Return to :
®
info@stateoftheart.it
©copyright STATE-OF-THE-ART®